PRO-DX EXCIEO: MODELS
Excieo H (GA2005H) - Hand Arm Vibration
Excieo W (GA2005W) - Whole Body Vibration Excieo C (GA2005C) - HAV and WBV Combined
Excieo E*(GA2005E) - Environmental & Engineering Vibration
Excieo M (GA2005M) - Multi-Function Vibration
APPLICABLE STANDARDS
ISO8041:1990 type 1 and latest draft of ISO8041 revision (Human
Vibration Meters) ISO8041: 2005
BS4675:1978, ISO2954:1975 Requirements for Instruments measuring
vibration severity ROHS compliant
ACCELEROMETERS
For Hand Arm Vibration: KD1006
Sensitivity: 10mV/g, Range (f.s.) ±500g,
Frequency response 2-5000Hz ±10%
For Whole Body Vibration: KD1007
Sensitivity: 100mV/g, Range (f.s.) ±50g,
Frequency response 0.5-3000Hz ±5%
For Environmental Vibration: TBA
For Engineering Vibration (single Axis): KD1003 Sensitivity: 100mV/g, Range (f.s.) ±80g,
Frequency response 2-13000Hz +5%
DISPLAY
Electro-Luminescent, Back-lit LCD panel (160x160 pixels)
Vibration Meter displays
Numerical: Parallel Tri-axial results
Bar graph: Time History plot
Real time clock: Day, Month, Year, Hour, Minute, Second language display:
English, French*, German*, Spanish* and Italian*
LEVEL RANGE
Linear Operating Range: 74 dB
HARM: RMS Range: 0.02-5000m/s2
Peak Range: 0.2-7000m/s2
WBV & Engineering: RMS Range: 0.001-500m/s2
Peak Range: 0.02-700m/s2
NOISE FLOOR
Typical: TBA
FREQUENCY WEIGHTING
Severity: Ws,Fa(0.4-100Hz), Fb(0.4-1250Hz), Fc(6.3-1250Hz)
HARM: Wh
WBV: Wb,Wc,Wd,We,Wj,Wk,Wm
FREQUENCY RANGE
1Hz - 6.3 kHz (electrical characteristics)
See accelerometer frequency range
DETECTORS
rms, rmq and Peak
REAL-TIME TRI-AXIAL MEASUREMENT
Real-time 3 channel measurement to give simultaneous display and
integration for tri-axial measurement.
MEASUREMENT PARAMETERS
All Variations (H, W, C, E & M)
Arms, Amin, Amax, Aeq, Amp, PEAK, Velocity, Displacement.
Variations W, C & M
CFmp, CF, VDV
Variations E & M
Fast-data, VDV
Units: m/s2, cm/s2 , g, dB, mm/s, µm
LOG TIME INTERVALS
1s -12hours user definable in one second intervals (all
parameters) up to 99 hours, cumulative every 12 hours.
TRIGGER FUNCTION
Trigger source: internal value (rising or falling)
TIMER FUNCTION
A Program Log Timer to give multiple Start-Stop timings (with sleep mode),
Repeat timer and programming function* to control multiple measurement
tasks (date independent). Real Time Clock and Calendar, plus measurement
duration.
REAL TIME CLOCK
Time and Date - accuracy better than 5s per day.
MEMORY
4Mb On-board FLASH, 2Mb available for data storage
INPUTS
3-Channel input for Voltage mode, charge mode (via interface unit) and Direct
DC Power input via external socket: 12V
SIGNAL OUTPUT
RS232: 19200 Baud, 8 bit, no parity, bi-directional, standard USB
interface, 'AC output
PROCESSING
A to D Converters: 20-bit Stereo, 16kHz sampling DSP: 5MIPS, 5MHz
processor Controller: 16bit, running at 16MHz
POWER REQUIREMENTS
12 volts DC (with mains adapter), Batteries: 6 x MN1500 (size AA),
Life: approximately 12 hours (alkaline batteries, continuous operation).
Simultaneous mains and battery operation.
SIZE AND WEIGHT
Dimensions: Height: 255mm (without Cable)
x Width: 100mm (max) x Depth: 49mm Weight: 780g approximately (including batteries)
MANUALS
Multi-language manuals*: English, French, German, Spanish and Italian.